EasyMax V2
|
EasyMax V2
Blood Glucose Meter
Meter Specifications |
Display |
LCD |
Data Transfer: |
Mini USB Port |
Batteries: |
2 x Alkaline 1.5V (AAA) |
Standards: |
IVD 98/79/EC
ISO 13485 : 2003
ISO 14971 : 2009
ISO 15197 : 2013
EN 980 : 2008
EN 61326-1 : 2006
EN 61326-2-6 : 2005
EN 61010-2-101 : 2002
|
Temp & Humidity: |
Operating : 10℃~40℃, below 85% R.H.
Meter storage : -10℃~65℃, below 90% R.H.
Strip storage : 2℃~30℃, below 90% R.H.
Transportation : 2℃~30℃, below 90% R.H.
|
Accuracy: |
95% within + 15 mg/dL when glucose <100mg/dL
95% within + 15% when glucose >100mg/dL
|
Precision: |
CV (correlation variation) is less than 5 % |
Error-Codes: |
Err : humidified / Used strips
LP : Low power, please replace new battery
001 with err icon : Not enough blood
005 with err icon : system fault
HI : test result is higher than 630 mg/dL
LO: test result is lower than 20 mg/dL
Ht : Temperature too high to have a test
Lt : Temperature too low to have a test
|
Capacity: |
480 results (glucose level, date & time) |
Blood Specimen: |
Capillary whole blood / Venous whole blood |
Sample Volume: |
0.6 ul |
Testing Time: |
5 Seconds |
Hematocrit range: |
30-55% |
Alternate Site Testing: |
Fingertip, Palm, Forearm |
Measuring Range: |
20 - 600 mg/dL |
Meter Dimension: |
83.7 x 51.2 x 22.9 mm |
Weight: |
44.6 g (without battery) |
Battery Life: |
Over 1,000 times |
Auto Shut-Off: |
Power off functions after 90 sec no key pressed |
Time Setting: |
12 Hours or 24 Hours (Optional) |
Principles: |
Electrochemical biosensor technology |
|
|
|