EasyMax EM Tag

EM Tag

Features
  • Meets ISO 15197:2013 accuracy requirement
  • Free from 61 substances
  • Under-fill blood volume detection
  • Before/after meal marking
  • No Coding
  • Personal Glucose Range Indicator

EasyMax TAG

Blood Glucose Meter

Meter Specifications
Display Liquid Crystal Display Module
Dimension : 28 x 29 mm
Data Transfer NFC
Batteries 2 x CR2032 Lithium Battery
High Quality > 2,000 Measurements
Standards IVD 98/79/EC
ISO 13485:2003
ISO 14971:2007
ISO 15197:2013
ISO 15223:2000
EN 980:2003
EN 60601-1:1990
EN 60601-1-2:2002
EN 61010-2-101:2002
Temperature and Humidity Operating: 10°C~40°C, Below 85% R.H.
Meter Storage: -10°C~65°C, Below 90% R.H.
Strip Storage: 2°C~30°C, Below 90% R.H.
Transportation: 2°C~30°C, Below 90% R.H.
Accuracy 95% within + 15 mg/dL when Glucose <100mg/dL
95% within + 15 mg/dl when Glucose >100mg/dL
Precision CV (correlation variation) is less than 5 %
Error Codes Err: humidified / Used strips
LP: Low Power, Please Replace Battery
001 with err icon: System Fault
002 with err icon: Not Enough Blood
HI: Test Result is High than 600 mg/dL
LO: Test Result is Lower than 20 mg/dL
Ht: Temperature Too High to Have a Test
Lt: Temperature Too Low to Have a Test
Capacity 880 Results (Glucose Level, Date & Time)
Blood specimen Capillary Whole Blood / Venous Whole Blood
Sample Volume 0.6 ul
Testing Time 5 seconds
Hematocrit Range 10~65%
Alternate Site Testing Fingertip, Palm, Forearm
Measuring Range 20~600 mg/dL
Meter dimension 74 x 47 x 12 mm
Weight 30 g (without batter)
Auto shut-off Power off functions after 90 seconds no key pressed
Time Setting 12 Hours or 24 Hours (Optional)
Principles Electrochemical Biosensor Technology
PC Connectivity NFC